These 200W, medical grade power supplies are designed with Skynet’s patented Ring-Free ZVS technology to ensure high efficiency and optimal size. This cool running series has no trouble operating at 70 degrees C ambient and has plenty of power in reserve to run at 250W continuously with only 10 CFM forced air cooling.
SNP-X20 series offers extremely high efficiency and is built with long-life components for enhanced reliability. This series features low leakage current and FCC & VDE class B EMI. Meanwhile, it complies with a wide variety of safety regulations so that it can be used in Medical, Industrial, Information Technology or other equivalent applications.
Features
• Both Medical & ITE Safety Approvals
• Active PFC
• Efficiency between 85% and 88%
• Operation from 0°C to 70°C by convection
• Only 1.5 inch height
• Remote Sense
• Safety Agency Approvals
UL 60950, UL 60601-1
CSA C22.2 No. 60950, CSA 601-1, CUL
TUV EN60950-1, EN60601-1
• EMI
FCC class “B”
EN55022 “B”
• EMS
EN61000-4-2,-3,-4,-5,-6,-11
• Active PFC
• Efficiency between 85% and 88%
• Operation from 0°C to 70°C by convection
• Only 1.5 inch height
• Remote Sense
• Safety Agency Approvals
UL 60950, UL 60601-1
CSA C22.2 No. 60950, CSA 601-1, CUL
TUV EN60950-1, EN60601-1
• EMI
FCC class “B”
EN55022 “B”
• EMS
EN61000-4-2,-3,-4,-5,-6,-11
Size
4 x 8 x 1.5 inch
102.3 x 201 x 38 mm
102.3 x 201 x 38 mm
Specifications
Input | VAC |
Voltage, V | 90 to 264 |
Frequency, Hz | 47 to 63 |
Current, A | inrush I <30A at 115VAC; <60A at 230VAC |
Power Factor | > .93 – Active PFC |
Harmonics | EN61000-3-2 class D |
Cooling | 200W Free Air Convection |
Power Good | N/A |
Holdup Time | >25 mS typical at rated load and 115VAC |
Transient | consult factory |
Efficiency | 85%~88% depending on model |
Shorted output | Protected auto recovery |
Over Load Protection | Auto-recovery |
Short Circuit Protection | Auto-recovery |
Over Voltage Protection | Latch-off |
EMS | EN61000-4-2, -3, -4, -5, -6, -11 |
Temperature | -20°C to 70°C; derating: 2.5% / °C > 50°C |
EMI | EN55022-B |
Safety Agency | UL 60950-1, UL 60601-1 CSA C22.2 No.60950-1, 601.1 EN 60950-1, 60601-1 |
DC Outputs
Output | 15V |
Rated Load | 16.5A /25A pk |
Regulation | ±1% |
Ripple | 100mV |