SNP-YL60 Medical 60W 3 x 5 inch
Series: SNP-YL6
SKU: SNP-YL60
Application: Medical & Health
Form Factor: Open Frame
Type: AC-DC
Output
No of Outputs: Outputs 4
O/P Voltage: 5V
, 12V
, 12V
, 5V
O/P Current: 3A
, 3A
, 0.3A
, 0.3A

These 60W, medical grade power supplies are designed with Skynet`s patented Ring-Free ZVS technology to ensure high efficiency and compact size.

The SNP-YL6 series offers up to 3.1 watts / in3 in a 3″ X 5″ X 1.18″ size with convection cooling.

The safety conformity of the SNP-YL6 series covers both IT and medical applications. SNP-YL6 series is built with long-life components for continued reliability.

This series features low leakage current and FCC & VDE class B EMI. Meanwhile, it complies with a wide variety of safety regulations so that it can be used in Medical, Industrial, Information Technology or other equivalent applications.

Features

• Only 1.18 inch height
• 3.1 Watt per cubic inch
• Both Medical & ITE Safety Approvals
• Efficiency between 75% to 86%
• Operation from 0°C to 70°C by convection
• Single side PCB for low assembly cost

Size

3 x 5 x 1.18 inch
76.2 x 127 x 30 mm

Specifications

Input VAC
Voltage, V 90 to 264
Frequency, Hz 47 to 63
Current, A inrush I <30A at 115VAC; <60A at 230VAC
Power Factor N/A
Harmonics N/A
Cooling Free Air Convection
Power Good N/A
Holdup Time 16 mS typical at rated load and 115VAC
Transient consult factory
Efficiency 75%~86% depending on model
Shorted output Protected auto recovery
Over Load Protection Auto-recovery
Short Circuit Protection Auto-recovery
Over Voltage Protection Latch-off
EMS EN61000-4-2, -3, -4, -5, -6, -8,-11
Temperature 0°C to 70°C; derating: 2.5% / °C > 50°C
EMI FCC-B EN55022-B, EN55011-B
Safety Agency UL 60950-1, UL 60601-1; CSA C22.2 No.60950-1, 601.1; EN 60950-1, 60601-1

DC Outputs

Output +5V +12V -12V -5V
Rated Load 3A / 5APK 3A / 5APK 0.3A / 1APK 0.3A / 1APK
Regulation ±3% ±3% ±5% ±5%
Ripple 50mV 120mV 120mV 50mV

Notes

1. At peak load, the output can last for 8 seconds without shut down.
2. The maximum combinational load of SNP-Y06E for +3.3V & +5V is 35W.
3. At factory, all outputs in 60% rated load condition, each output is checked to be within the accuracy range while the main output is setting to within the specified accuracy range at rated load.
4. Line regulation is defined by changing ±10% of input voltage from nominal line at rated load.
5. Load regulation is defined by changing ±40% of measured output load from 60% rated load at another output set to 60% rated load.
6. Ripple & noise is measured by using 15MHz bandwidth limited oscilloscope and terminated each output with a 0.47uF capacitor at rated load and nominal line.
7. Hold up time is measured from the end of the last charging pulse to the time which the main output drop down to regulation limit at rated load and nominal line.

Mechanical